Note : Pre-order item
Model : ST-105L
Class 1 Integrating Sound Analyzer Meter
Class 1 comply with IEC – 61672-2013 Class 1
IEC – 61260-1995 Class 1
Features :-
- Statistic analysis
- 24 hr measurement
- Integral measurement
- Real time analyzer in 1/1 and 1/3 octave band
- It simultaneously measures A, C, Z and Fast, Slow, Impulse
- Real time clock with calendar
- Sampling frequency : 20.8 µs (48 KHz)
- Wide frequency range : 10 Hz – 20 KHz
- Wide measurement range : 25 dB – 140 dB
- USB and RS-232 interface
- Wide dynamic range : 90 dB
- Fast, Slow, Impulse, PeakC+, PeakC-
- Measurement function : Lxyp, LAFmax, LAF5, LAF10, LAF50, LAF90, LAF95, LAFmin SD, LAeq1s, LAeqT, LAE, LAfeqT, LD, LN, LDN, Lxyi, Lxeq1s, LxeqT, E, Cpeak+, Cpeak-, LAFmin, LAFeqT, LASeqT, LAIeqT, Lfmeq1s, LfmeqT
Analysis frequency range :
1/3 Octave ( 34 bands )
|
10 Hz (ST-105D), 12.5 Hz, 16 Hz, 20 Hz, 25 Hz, 31.5 Hz, 40 Hz, 50 Hz, 63 Hz, 80 Hz, 100 Hz, 125 Hz, 160 Hz, 200 Hz, 250 Hz, 315 Hz, 400 Hz, 500 Hz, 630 Hz, 800 Hz, 1 KHz, 1.25 KHz, 1.6 KHz, 2 KHz, 2.5 KHz, 3.15 KHz, 4 KHz, 5 KHz, 6.3 KHz, 8 KHz, 10 KHz, 12.5 KHz, 16 KHz, 20 KHz (ST-105D)
|
Parameters and performance of Octave and 1/3 octave :
Function
|
Noise’s real-time octave and 1/3 octave spectral analysis and integrating measurement
|
Frequency weighting
|
Parallel (simultaneous) A, C, Z. The spectral analysis interface is composed by the frequency spectrum. The total analysis interface is realized by digital filtering
|
Filter type
|
Parallel (real-time) octave and 1/3 octave band filter, G=2, digital filter. The noise exposure (E)’s measuring range is 0 – 65.535 Pa2h
|
Specifications :
Measurement parameter
|
Lfmeq1s , LfmeqT , Lxyp, Lxyi , Lxeq1s , LxeqT, LAE, E, Cpeak
|
Applications
|
Machine’s noise analyses
Noise spectrum analyzer
|
Display
|
LCD with back light (240 x 160 dots)
|
Integrating
|
Yes
|
1/3 OCT
|
Yes
|
128 records storage
|
Yes
|
IEC 61672-1:2002 Class 1
ANSI Si.4 Type 1
CNS No. 7129 Type 1
JJG-188:2002 Class 1
|
Yes
|
IEC 61260-1995 Class 1
|
Yes
|
Display refresh
|
10 Hz for value; 1 Hz for graph
|
Microphone
|
½” pre-polarized condenser microphone build in preamplifier
Sensitivity : 50 mV/Pa
Frequency range : 10 Hz – 20 KHz
Heat noise : < 16 dB(A)
|
Measurement range
|
25 dB – 140 dB (A)
30 dB – 140 dB (C)
35 dB – 140 dB (Z)
|
Dynamic range
|
> 90 dB
|
Instrument background noise
|
< 13 dB (A), 15 dB (C), 25 dB (Z)
|
Maximum Peak C sound level measurement
|
50 dB – 143 dB
|
Time weighting
|
Fast, Slow, Impulse, PeakC+, PeakC-
|
Frequency weighting
|
A / C / Z
|
Integrating time
|
1 second – 24 hour, set in a given range or randomly
|
Frequency range
|
10 Hz – 20 KHz
|
Range gain
|
- 10 dB, 0 dB, 10 dB, 20 dB, 30 dB, 40 dB
|
Range error
|
|
Self-generated noise voltage
|
< 4µV (1 Hz – 23 KHz)
|
Measuring voltage range
|
15 µV – 10 Vrms
|
Sampling frequency
|
20.8 µs (48 KHz)
|
Analog output
|
A C
|
Starting time
|
< 10 seconds
|
Interface
|
USB interface, mini B type, complying with USB 1.1, compatible with USB 2.0 which can transfer the measuring result to PC and transfer the data to the flash disk up to 8 G
|
Power
|
1.5V x 4 alkaline batteries
|
External power
|
AC 100 – 240V to DC 5V
|
Dimension
|
285 x 90 x 39 mm (L x W x H)
|
Weight
|
500g including batteries; 2.5 kgs including packaging
|
Note : All the above specifications are subject to change without prior notice
Country of origin : Taiwan.
Warranty : 1 year against manufacturing defects. Remarks : For warranty claim :- 1) There is NO one to one exchange. 2) Unit need to sent back to manufacturer for rectification & repair.
Delivery : 3 to 5 days (in-stock unit) else 2 to 4 weeks (pre-order unit) subject to factory confirmation
Free Shipping Within Malaysia
Tel : +6012 728 1063
Email : enquiry@giitechnology.com |